GasAlert Extreme
Calibration and Setting Alarm Setpoints
Set Span
The display shows the current calibration gas setting.
The display flashes prompting you to input a new calibration
gas concentration, or you can accept the current value for the
span (sensitivity adjustment).
4. To change the calibration gas setting, press D or
E until the display matches the concentration of
the calibration gas.
5. Press C to save the new value.
Note
If you do not press any pushbuttons within 10
seconds of viewing this display, the detector
automatically saves the displayed concentration of
the calibration gas.
If you change the calibration gas concentration but
pause for 10 seconds before pressing C, the
detector rejects the new value. The display shows
NO, the audible alarm beeps six times, and the
detector retains the original value.