GasAlert Extreme
User Manual
Note
Verify that the calibration gas being used matches
the span concentration values that are defined for
the detector. For more information, refer to Span
Concentration.
The set span screen displays a flashing gas
cylinder.
Note
The flashing gas cylinder icon does not display for
oxygen (O
2
) detectors.
6. Apply the calibration gas.
7. Apply gas to the sensor at a flow rate of 500
ml/min.
(for NH
3
, Cl
2
, and ETO: 1000 ml/min.)
The gas readings change as gas is applied to the
sensor. When the detector senses a sufficient
concentration of gas (approximately 30 seconds), the
audible alarm beeps once.
The detector then begins spanning the sensor as follows:
NH3, Cl2, ClO2, O3, and ETO: 5 minutes;
O2: 30 seconds;
other gases: 2 minutes (approximately).
The audible alarm beeps three times when the span is
complete.
Successful Span
If the span is successful, the LCD automatically displays the
calibration due date screen.
Unsuccessful Span
If the detector fails to span a sensor successfully, the LCD
displays an error message.
The detector flashes, vibrates, and emits a long tone and
then automatically proceeds to the alarm setpoint screens.
44